ASTM F1195
$35.00
Standard Specification for Rigid Laryngoscopes for Tracheal Intubation-Hook on Fittings for Fiberilluminated Blades and Handles (Withdrawn 2008)
Published by | Publication Date | Number of Pages |
ASTM | 07/10/1999 | 5 |
Description
ASTM F1195 – Standard Specification for Rigid Laryngoscopes for Tracheal Intubation-Hook on Fittings for Fiberilluminated Blades and Handles (Withdrawn 2008)
1.1 The purpose of this specification is to ensure that reusable, detachable fiber-illuminated hook-on blade and handle combinations that engage shall lock and illuminate when in the operating position.
1.2 Dimensions and tolerances of fiber-illuminated systems shall be provided.
1.3 This specification provides methods of identifying by color the various fiber-illuminated systems and compatible blades and handles within the system.
1.4 This specification covers reusable handles and blades. It does not cover disposable components or systems.
1.5 Nonfiber-illuminated blades and handles with the lamp in the blade shall conform to Specification F965.
Product Details
- Published:
- 07/10/1999
- Number of Pages:
- 5
- File Size:
- 1 file , 87 KB
- Note:
- This product is unavailable in Russia, Ukraine, Belarus